Abstract
OBJECTIVE: To assess the agreement between artificial intelligence (AI) interpretation and expert emergency physician sonologist review of echocardiograms performed in the Emergency Department (ED).
METHODS: Retrospective multicentre agreement study in four Australian EDs. A total of 195 echocardiograms (2016-2023) underwent three-stage adjudicated human expert review. Human reports were compared with us2.ai reports. Agreement across 10 validated parameters was reported as percentage exact match with 95% confidence intervals and kappa under two pre-specified tiers: Guideline Compliant (GC; ASE/EACVI multi-view criteria) and Pragmatic (PR; any available measurement). Reference standard reliability was quantified.
RESULTS: Pooled agreement was 76% at both tiers (GC n = 875; PR n = 1323); AI returned an interpretation in 46% of evaluations at GC and 70% at PR. Four structural parameters reached 78%-87%: right atrial size 87% (κ = 0.62), right ventricular (RV) contractility 83% (κ = 0.42), left ventricular (LV) size 83% (κ = 0.52) and RV size 78% (κ = 0.37). Binary LV systolic agreement was 79% GC and 83% PR. Expert reviewers agreed on 85% of interpretations. Pericardial effusion, aortic regurgitation and mitral stenosis, present in 7%-20% of patients, are not reported in the current us2.ai algorithm.
CONCLUSION: AI agreed with expert review in about three-quarters of evaluations but provided a guideline-compliant interpretation in less than half; a pragmatic approach raised yield to 70% at the same agreement. The AI withheld classifications that experts could reach; its validated scope covers only part of what ED echocardiography must address, and reports did not mark unassessed parameters. This study reports agreement, not accuracy. For now, human expert review of AI reports is required.
METHODS: Retrospective multicentre agreement study in four Australian EDs. A total of 195 echocardiograms (2016-2023) underwent three-stage adjudicated human expert review. Human reports were compared with us2.ai reports. Agreement across 10 validated parameters was reported as percentage exact match with 95% confidence intervals and kappa under two pre-specified tiers: Guideline Compliant (GC; ASE/EACVI multi-view criteria) and Pragmatic (PR; any available measurement). Reference standard reliability was quantified.
RESULTS: Pooled agreement was 76% at both tiers (GC n = 875; PR n = 1323); AI returned an interpretation in 46% of evaluations at GC and 70% at PR. Four structural parameters reached 78%-87%: right atrial size 87% (κ = 0.62), right ventricular (RV) contractility 83% (κ = 0.42), left ventricular (LV) size 83% (κ = 0.52) and RV size 78% (κ = 0.37). Binary LV systolic agreement was 79% GC and 83% PR. Expert reviewers agreed on 85% of interpretations. Pericardial effusion, aortic regurgitation and mitral stenosis, present in 7%-20% of patients, are not reported in the current us2.ai algorithm.
CONCLUSION: AI agreed with expert review in about three-quarters of evaluations but provided a guideline-compliant interpretation in less than half; a pragmatic approach raised yield to 70% at the same agreement. The AI withheld classifications that experts could reach; its validated scope covers only part of what ED echocardiography must address, and reports did not mark unassessed parameters. This study reports agreement, not accuracy. For now, human expert review of AI reports is required.
Publication information
Emerg Med Australas. 2026 Oct 10;38(5):e70364. doi: 10.1111/1742-6723.70364
File(s)![Thumbnail Image]()
Loading...
Name
Emerg Medicine Australasia - 2026 - Manivel - Artificial Intelligence Echocardiographic Interpretation in the Emergency.pdf
Description
© 2026 The Author(s). Emergency Medicine Australasia published by John Wiley & Sons Australia, Ltd on behalf of Australasian College for Emergency Medicine.
This is an open access article under the terms of the Creative Commons Attribution-NonCommercial-NoDerivs License, which permits use and distribution in any medium, provided the original work is properly cited, the use is non-commercial and no modifications or adaptations are made.
Size
1.32 MB
Format
Adobe PDF
Checksum
(MD5):b2e824679c077a8ed23c8f3975c6d6cc
Date Issued
2026-10-10
Type
Journal Article
Journal Title
Emergency medicine Australasia : EMA
Permanent link to this record
ISSN
1742-6723
Journal Volume Number
38
Pages
e70364
Owning collection